1. OVERVIEW

    WinCal 3.2 - Cascade Microtech VNA Calibration Software.

    This file contains information that either became available too late for
    manual and help file publication or is not appropriate for those mediums.
    
    
2. SUPPORTED SYSTEMS

    WinCal 3.2 has been tested and certified to work properly with the following
    software, drivers, and operating systems.  Serious problems may occur when
    using versions prior to the ones listed here.

    Probers: 
      Manual probers
      Nucleus v2.5.4   (or later 2.5 version)
      Nucleus v2.6.1   (or later 2.6 version)
      Nucleus v2.7.0   (or later 2.7 version)
      PS-21   v2.0.13  (or later 2.0 version)*
      Note: PCS-based probers are NOT supported

      * Prober controller service pack and internet explorer upgrade also 
        required.

    Important!
      WinCal 3.2 will NOT run with Nucleus or PS-21 versions older than 
      those listed above.  Contact Cascade Microtech Service (see manual) for 
      information about obtaining a free upgrade of the prober software.
      
    Third Party GPIB and VISA (required, see section 5 below):
      National Instruments GPIB driver v1.6  (or later version)
         (Please refer to <http://www.ni.com/support/gpib/versions.htm> 
          to see if your GPIB version is compatible with your hardware 
          and operating system.)
      National Instruments VISA runtime v2.5 (or later version)
        (or)
      Agilent I/O library version M with VISA included (or later version)
      
    Third Party Other:
      Sentinel System Drivers v5.39.2 or later
                 
    Operating Systems:
      Windows 98 Second Edition
      Windows NT 4.0 (Service Pack 6a) Note change from SP6 to SP6a (see below)
      Windows 2000 (Service Pack 3)
      Windows XP (Service Pack 1)
      Note: Windows 3.1, 95 and ME are NOT supported
		
		Other System dependencies:
			For the new HTML-based On-Line Help to work, Internet Explorer 5.0 
			or later needs to be installed.	This will probably not be an issue for
			Windows 2000 or XP, but could be a problem with older NT 4.0 systems.
			Microsoft no longer offer downloads of older versions of Internet Explorer.
			Only the current Internet Explorer 6.0 (IE 6.0)is available at their website.
			However, IE 6.0 requires Service Pack 6a (SP6a) on NT 4.0, but many probers only 
			have SP6 (without the "a" fix).  It is possible to download SP6a for NT4, 
			install it, and then download and install IE 6.0.  If you have IE 5.0 or IE 5.5
			available that would be enough as well.  
			Contact Service to get directions as needed.

             
3. NEW/FIXED IN VERSION 3.2
   
   NEW:
     - Support for Rohde & Schwarz VNA models ZVK and ZVM (four sampler models)
     - Some corrections to make the PS300 prober driver work correctly
     - The step sizes of wide pitch ISS's were corrected in the wincal.ini file
     - A problem with PNA calkit download was fixed
     - All additions and correcions in WinCal 3.1.2 and older 
       (For changes in previous versions see the end of this file)


4. INSTALLATION - GENERAL

    WinCal can either be installed to work with an existing Nucleus (2.0 or 
    later) based prober, on a PS21, or to run stand alone and work with a 
    manual station.  

    Nucleus/PS21 should be installed first to enable setting semi-automatic 
    stations as the default.  However, this can be changed manually later.

    We strongly recommend that WinCal is installed in a separate directory from 
    Nucleus or PS21.
                          

5. INSTALLATION - VISA and GPIB

    In order to support different GPIB cards WinCal now requires
    that the VISA (Virtual Instrument System Architecture) interface
    is installed from the vendor of the card.  The VISA software comes
    with the CD for the cards, but can also be obtained from the vendors 
    web site (free).  This may be needed to get a late enough version. 
    Please refer to section 2 (above) for supported driver versions.
    This does not apply to demo installations without any GPIB boards, 
    where no GPIB drivers are needed. In this case you don't need to install
    any GPIB driver software.

    If Using National Instruments GPIB Board:
    -----------------------------------------
    A GPIB board and it's driver from National Instruments should be installed 
    before WinCal is installed. If you have a Summit, S300 or PS21 probe station
    this was done at the factory.  Otherwise use National Instruments 
    instructions to install the GPIB drivers and the runtime library for VISA.  
    For the latest news and final word of GPIB/VISA compatibility see National
    Instruments' web site at http://www.ni.com/support/gpib/versions.htm
    
    If Using Agilent GPIB Board:
    ----------------------------
    A GPIB board and it's driver from Agilent should be installed 
    before WinCal is installed.  The default choice installs the VISA software 
    as well as the lower level GPIB driver.


6. INSTALLATION - HARDWARE KEY DRIVERS  (N/A for demo)

    The hardware key (Sentinel SuperPro) is attached to 
    the parallel port on the PC. Other Windows programs 
    may also access that port, the printer spooler being 
    the obvious one. The regular WinCal installation 
    program adds a parallel port driver.  


7. NUCLEUS and GPIB control

   Using an external measurement computer via a GPIB switch box
   ------------------------------------------------------------
   A probe station with Nucleus or PS21 software can be controlled by en external 
   computer via GPIB commands.  If WinCal is installed on such a probe station, then
   a GPIB Switch box (see appendix in the WinCal manual) can be used to switch between WinCal 
   controlling the VNA and the external computer controlling the VNA.  The Nucleus hardware 
   setup dialog should be used to add the component which enables it to be controlled by an 
   external computer via GPIB.  WinCal and the switch box makes it work by taking turns being GPIB
   system controller.  
   
   Using an external measurement computer via "Nucleus GPIB Driver" PC software
   ----------------------------------------------------------------------------
   If the other (external) computer is a Windows based PC, the new software tool 
   "Nucleus GPIB Driver" (see separate folder on CD and appendix in the manual) could be used.  
   In this case WinCal will be installed on the external PC and send commands to Nucleus via GPIB.
   Nucleus GPIB Driver software will also be installed on the external PC and will act as a 
   "simulated" Nucleus to WinCal.  It will take the commands and pass them on via GPIB to Nucleus
   on the Probe Station PC.  Here you also need to configure Nucleus hardware setup dialog to support
   GPIB remote control.  
   
   By default, Nucleus GPIB Driver will add the following entries to "WIN.INI":
   
     [CascadeMicrotech]
     SysMgrVersion=2.0
     SysMgrWindowCaption=simulator
   
   If either of these entries are missing or incorrect, please enter the correct values directly in "WIN.INI"
   and restart WinCal and Nucleus GPIB Driver software.
   
   Nucleus GPIB Driver Limitations:
      - After installing Nucleus GPIB Driver, each time WinCal is started the Nucleus GPIB Driver 
        will also be started.  To discontinue using Nucleus GPIB Driver, please uninstall the 
        software.
      - Nucleus GPIB Driver software does not support Agilent GPIB cards.
   
   Using the Probe Station as measurement computer
   -----------------------------------------------
   If, on the other hand the measurement software is also running on the probe station (e.g. Wavevue),
   the Nucleus hardware setup dialog should be used to remove the component which enables it to be 
   controlled by an external computer via GPIB.  In this case there is the no need to switch GPIB
   control, since all controlling software runs on the prober, and there is no external computer 
   controlling Nucleus.  The Probe Station will NOT be controlled via GPIB, only by internal 
   communications (COM Automation)from WinCal and any other Automation application on the same PC.
   

8. STANDARD USE OF WIN.INI 

   How WinCal and Nucleus/PS21 find each other
   -------------------------------------------
   WinCal uses an entry in its win.ini file to find the location of Nucleus or 
   PS21 software.  The key names are SysMgrPath and SysMgrVersion.  See 
   examples below.
   
   The Nucleus toolbar can be configured with a launch button for WinCal.  See 
   the Nucleus User Manual for more information.

   If no prober entry at all is found in the win.ini file, then WinCal will 
   configure itself to run with a manual prober and not allow any automatic
   moves.

   Example Nucleus System Manager case:

     [CascadeMicrotech]
     SysMgrPath=C:\PROGRA~1\Cascade\Nucleus\SysBin\
     SysMgrVersion=2.6
     RemoteThetaEnabled=0
     WinCalPath=C:\Program Files\Cascade\Wincal3_2\ 
     WinCalVersion=3.2.0
     GPIBLogToFile=0

   Example PS21 System Manager case:

     [CascadeMicrotech]
     SysMgrPath=C:\RAVEN\BIN\
     SysMgrVersion=2.0
     RemoteThetaEnabled=1
     WinCalVersion=3.2.0
     WinCalPath=C:\Program Files\Cascade\Wincal3_2\ 
     RemoteThetaEnabled=1
     GPIBLogToFile=0

   Important Note:  The Win.ini file will include different entries if the 
                    Nucleus GPIB Driver for WinCal is installed.  Please see 
                    section 7 for details on Win.ini entries for that scenario.

9. INI FILE SETTINGS FOR SOME SPECIAL CASES

  In the WINCAL.INI file, which is installed to the same folder as WinCal itself
  
  There are several settings in the WINCAL.INI file 
  The keys that are of interest to an end user are documented in the file itself,
  but most keys are used only by the program and should not be changed by hand.

  In the [CascadeMicrotech] section of the WIN.INI file (in the Windows folder):
    
      AlwaysAddLF=1 
        If added, the Line Feed character is always added to output
        strings.  Try it for older Agilent VNA firmware if you have problems,
        but do not use it for any Anritsu/Wiltron VNA.  They get LF anyway.

      GPIBLogToFile=1
        If added, a log of all GPIB action from program start is
        written to a file named wcgpib.txt in the WinCal directory.
        Do not leave this on.  It creates a log file that can grow very large
        and also slows down operations.  Use only for trouble shooting!
    
      RemoteThetaEnabled=1
        Use with PS21 version 2.0 or later to enable remote theta commands. 
        Must be zero (=0) for any non-motorized-theta prober.  Adds theta to
        recorded positions.
        Note1: The S300 has movable theta, but the auxiliary chucks do not move 
        with the wafer chuck, so WinCal does not need to move theta.
        Note2:  There may be new station models with movable theta and auxilliary
        chucks that moves with theta (like a PS21). If so, you should set this 
        switch to 1, or remove it and let WinCal detect theta capability by 
        calling Nucleus.
        
      AlsoSaveRawInTouchstoneFiles=1
        When a raw calibration file is saved, this would make WinCal also save a 
        regular Touchstone format (.s1p & .s2p) text file with the raw trace reading 
        of the standards used for calibration.  
        File names are: 
        RawOp11.s1p, RawOp22.s1p, RawLo11.s1p, RawLo22.s1p, RawSh11.s1p, RawSh22.s1p,
        RawThru.s2p, RawThA12.s1p, RawThA21.s1p, 
        RawLineN.s2p, RawLineNA12.s1p, RawLineNA21.s1p, where N is the line number

      SysMgrWindowCaption="WinCal Nucleus-GPIB Driver"
        Only for special use with "WinCal Nucleus-GPIB Driver".  A way to override
        the default windows caption WinCal uses to see the Nucleus/SysMgr is running.
        Do not use this for normal operation on the prober.

    
  In the [Configuration] section of the DRVNA.INI file:

      VNADebugEcho=0
        If changed to 1, the VNAs that support it will show incoming GPIB 
        commands on their display

      W360VersionCheck=0   
        If added, the version check for Wiltron 360 firmware is skipped.
        Beware: There is no guarantee that older firmware will work correctly,
                it is expected to have problems.

      IgnoreVnaErrMsg=1
        If added, then caution and error messages on the Agilent VNA's will NOT
        be considered a problem by WinCal. This can get rid of some "false" 
        error conditions that would stop WinCal from accepting a measurement as 
        valid, but it is then the operators responsibility to judge if it is
        OK to continue operations anyway.
        Note:
        If you get many phase lock errors, talk to Agilent about a firmware upgrade.


10. VNA LIMITATIONS

   - When sending the calibration kit to the VNA, some values from the Calibration 
     Definitions dialog are not sent to the VNA's calibration kit.  
     These values are:
       -> Open offset
       -> Short offset
       -> Load resistance
       -> Thru offset loss
       -> Thru offset loss frequency
       -> System impedance

   - Log Freq sweep type is not supported by any instrument driver.

   - Some specifics for the Agilent PNA family of VNA's:
       a) Cal sets will be saved to a file named "CalSetX.cal", where "X" is 
          the cal set number selected in the Calibration Setup dialog.    
          Instrument states will be saved to a file name "InstrumentStateY.cst",
          where   "Y" is the instrument state number selected in the Calibration 
          Setup   dialog. Both files can be located in the following directory on 
          the PNA:  
          "C:\Program Files\Agilent\Network Analyzer\Documents".
       b) When WinCal saves a cal set to file, an entry is appended to the 
          "Select A Cal Set" dialog box on the PNA.  The number of this new 
          entry does not match the cal set number selected in the Calibration 
          Setup dialog.  In addition, the cal set type listed in the "Select A 
          Cal Set" dialog box may not match the type of calibration performed by 
          WinCal.
       c) Firmware version A.03.52 or later is required for the PNA.

   - Some specifics for the Anritsu VNA's:
     - Do not combine low IF bandwidth with using high (or any) averaging for 
       Anritsu VNA's.  It makes them very slow and timeouts are likely to occur. 
       This is essentially true for any VNA, including Agilent models.
     - The Wiltron 360 (now Anritsu) VNA does not support download of a SOLT 
       calkit.
     - Firmware version 1 or later is required for the Agilent 8703  
       The 8703A has communication timing problems and you may have to add 
       GPIB delays in that case.  The 8703B model has no such problems.
     - Firmware version 4.02 or later is required for the Anritsu/Wiltron 360B
     - Firmware version 2.03 or later is required for the Anritsu 27xxx
     - The oldest firmware version Anritsu 37xxx was tested with is 4.01.  
       No earlier version is supported, but it may work with version 2.03 or later.

   - Some specifics for the Rohde & Schwarz ZVK and ZVM:
     - Minimum required firmware version is 3.52
     - This instrument requieres a "fake cal" to accept external error correction
       terms.  That is why you will see a fast "cal" go by after WinCal has read the raw 
       data it needs. This "fake cal" is done at fastest possible sweep speed (max IF) and if 
       the IF setting later is lower than 300 Hz you will see "CA?" as an indication on the VNA
       screen of a cal being ON, as opposed to the normal "CAL".  This is not a problem. 
       The cal is still valid.  To always get the "CAL" indication you can turn off making 
       the fake cal at max IF, but then you will have to wait for WinCal's readings and the
       fake cal readings at the same IF setting (typically low sweep speed for stability)
       You can turn off fast "fake cal" by adding RohdeSchwarzZVFastDummyCal=0 in the 
       [Configuration] section in drvna.ini
     - The ZVx has no calkit term for load inductance, so that value is not sent with a WinCal
       probe calkit.  This only affects front panel SOLT calibrations on te VNA.  When WinCal 
       performs the SOLT calibration, it is using this value.
       
       See also "CRN-WinCal 3.2- Rohde & Schwarz VNA Specifics.txt" in the "document\tech notes" folder



11. OPERATING SYSTEM LIMITATIONS
   - Cascade Microtech recommends that computers running WinCal on Windows 2000 
     be configured to run in the "standard" mode. If you are using a computer 
     with Windows 2000 (purchased from Cascade Microtech) your computer is 
     already running in the standard mode. Microsoft issued support notes 
     Q266344 and Q299877 indicating that computers running Windows 2000 in the 
     "ACPI" mode are susceptible blue screening or unexpected system wide hangs.
     If you experience these problems with your computer, please contact your 
     internal IS or computer support group for help to return your system to the
     "standard" mode.  These modes relates to PC hardware/BIOS power management.


12. USE OF MOTORIZED POSITIONERS

   - The motorized positioners are handled by Nucleus.  WinCal is limited to 
     recording and restoring positions (both stage and positioners) through
     Nucleus remote commands.

     If the positioners were set manually by turning the knobs, be quick to 
     record the desired position once, because the positioners have a "hold
     position" feature for small manual moves, to avoid vibration problems. 
     It stops after the first record in WinCal, and it is then possible to 
     fine adjust probe overtravel etc. by hand.

   - Motorized positioners are not supported by PS21 (version 2.0)


13. USE OF ZONES IN NUCLEUS 2.1 (and later) 

    Probe Stations running Nucleus have a concept called zones. The center wafer
    chuck is one zone and the auxiliary corner chucks are other zones. Each zone
    has it's own "Z contact" and "Z separate" setting in Nucleus.
    When WinCal is moving to a recorded position on an ISS in one of the aux- 
    zones a Nucleus command is used that will find the lowest "Z separate" value 
    (Z-coordinate) of all zones crossed (if more than one).  The end Z value
    for WinCal moves is always recorded in WinCal itself, but it is still 
    required to set the "Z contact" and Z separate" values in Nucleus for all 
    applicable zones in order to get correct clearance when moving.
        
    WARNING:
    The concept of zones relies on the probes being in the exact center, since
    the wafer zone is defined as a radius from the center. If the probes are 
    off center, then the only safe way to move is to have "Z separate" settings
    that results in a lower absolute Z values than any top surface ("Z contact")


14. FILES
  
  WinCal files 

    Code files:

    WINCAL.EXE    - Main program
    WGPIB32.DLL   - WinCal's interface to National Instruments GPIB.DLL
    WTOOL32.DLL   - Misc utilities used by WinCal
    DRVNA32.DLL   - VNA drivers
    MALRM32.DLL   - The LRM/LRRM calibration
    MASOL32.DLL   - The SOLT/SOLR calibration
    MATRL32.DLL   - The TRL calibration

    INI files:

    DRVNA.INI    - Initialization for the VNA drivers (DRVNA.DLL)
    WINCAL.INI   - Initialization for WinCal
    PROBES.INI   - Initialization of probe specific data

    Data files and misc. others:

    *.S1P        - Touchstone 1-port S-parameter files (plain S1P format)
    *.S2P        - Touchstone 2-port S-parameter files (plain S2P format)
    *.MDF        - Touchstone 2-port S-parameter files (MDIF format)
                   Some are used by the virtual VNA driver for PPR simulation.

                   Used by virtual VNA driver:
    VIRTUAL.SOL  - Binary file with a sample of raw SOLT/SOLR calibration data
    VIRTUAL.LRM  - Binary file with a sample of raw LRM/LRRM calibration data
    VIRTUAL.TRL  - Binary file with a sample of raw TRL calibration data
    VIRTUAL.PTS  - Binary file with a sample of raw Probe Test data

    WCREADME.TXT - This file.  Misc. other text files are also installed.
    WINCAL31.HLP - On-Line Help for WinCal

  Third party files installed by WinCal: 

      In windows\system dir:
        MSHFLXGD.OCX  - MS FlexGrid Control (only if not already there and VB 
                        sample selected)
        
      In target WinCal dir:
    PEGRP32A.DLL - GigaSoft 32-bit Graphics tool
    PEGRAPHS.HLP - GigaSoft Graphics tool help
    CW3230.DLL   - Borlands 32-bit runtime library, version 5.0  
    BDS52f.DLL   - Borlands 32-bit class library, version 5.02  
    OWL52f.DLL   - Borlands 32-bit ObjectWindows library, version 5.02  


15. CHANGE HISTORY
==================

VERSION 3.1.2
   
     - Support reading up to 6401 points on the PNA (was 1601 max)
     - Support for the 110 GHz Infinity GSG probe was added.
     - The latest On-Line Help file is included.  No need to download a correction.
     - In PPR, if the measured H21 trace didn't cross zero, a graph error was 
       generated.  Now the graph displays and it is clear what happened.
     - A problem displaying a perfect OPEN (1,0) in the Smith chart was fixed
     - Stepping sizes beyond 1270 microns are no longer ignored when reading 
       wincal.ini for ISS specs. 
     - The stepping size for ISS P/N 109-531 was corrected to be 1800 microns (was 1500)
     - Reading 1-port corrected data with a PNA in now working correctly.  This would have
       affected Probe Test for example.  
     - Some smaller issues with history traces in the graph were corrected.  Did not always
       recognize the need to erase history and start over.
  
VERSION 3.1.1
   - Added new software tool "Nucleus GPIB Driver" to this release.  
     See Section named "NUCLEUS and GPIB control" for more details.
   - New probe calibration coefficients added for the Infinity Probe probes.
     The Special probe selection now keeps it values instead of being auto calculated
     to defaults each time. (This applies to SOLT calibration and Probe Test)
   - Improved safety: a warning is now displayed if Nucleus contact height
     setting has changed since a WinCal position was recorded.
   - Support was added for the Anritsu Scorpion (46xxx -series) VNA
   - Ability to send SOLT calkits to both Anritsu Lightning (27xxx & 37xxx) and 
     Scorpion (46xxx) series VNA was added.
   - New on-line help created for the newer standard HTML based Windows help engine.  
     Also updated content and added more screen shots.
   - Various other minor bug fixes and enhancements
   
   FIXED:
    (Misc fixes)
   - Old data does not persist when Overlay Traces feature turned off
   - Old data does not persist when overlaying traces with different frequency
     sets
   - If another remote program sets Nucleus units to mils, WinCal will change it 
     to microns for it's own use, and then restore to mils when done.  N/A for PS21 and PS300
   
    (PNA fixes)
   - SOL calibrations now work on the PNA
   - Values correctly loaded for PNA Calibration Kit
   - Multiple calibrations will no longer crash the PNA
   - Port extensions now applied to S-Parameter measurements

VERSION 3.1

  Enhancements and corrections
  ----------------------------
   - Support for Agilent PNA Network Analyzers 835x series and 
     836x series added.
   
   - Graphical overlay of traces in S-Parameter tool added, as well as the 
     ability to save S-parameter files in Magnitude-Angle and dB-Angle 
     (in addition to real-imaginary).

   - Support for Agilent GPIB added.

VERSION 3.0

  Enhancements and corrections
  ----------------------------
   - WinCal 3.0 is a 32-bit application. Long file names are now supported  
     and it is more stable on Windows NT and 2000
   - For probe definition in SOLT and Probe Test: If the pitch 250 microns
     was selected, the code assumed the ISS for 250 microns or less pitch.
     Now there is a choice, since 250 micron probes can be used on both narrow
     and wider type ISS:s
   - Two new (optional) keys have been added to the win.ini file as follows:
     ...
     [CascadeMicrotech]
     GpibReadDelay=0
     GpibWriteDelay=0
     ...
     These two keys allow GPIB access to older VNAs which may require longer
     communication delays due to increasing hardware speed on the PC end.
     Delays are listed in milliseconds.
   - Communication with the VNA 8703A was stabilized by adding a small delay 
     after each command sent.  This got rid of strange communication hang-ups.
     The delay is 300 ms by default, but that can be changed in the file 
     DRVNA.INI, using the key CmdDelayMs8703 in the [Configuration] section.
     Note: the 8703B has much later and better firmware and the delay is not 
     needed for this model.  Set it to 0 if you have the B model.
   - A phase jump problem with the SOLR calibration was fixed
   - For Nucleus 2.1 the zones concept is used. (See section 11 below)

VERSION 2.31
 
  Enhancements and corrections
  ----------------------------
  - Fixed problem with recognizing option 400 on VNA 87xx later than rev. D
  - Updated to the latest version of vendor drivers for the hardware key 
  - Added more product information document files on distribution CD
  - Default Start Menu on NT is now in the "all users" section (same as Nucleus)
  - Fixed a problem with using software align for a custom ISS when X-distance=0
  - Added win.ini switch AlsoSaveRawInTouchstoneFiles=0/1 to turn off/on 
    saving of raw cal data to S1P and S2P files when saving raw cal file.
  - Fixed GPIB switch box problem when using "ToVNA" in Probe Setup Dialog

VERSION 2.3   

  Features
  --------
  - A PS21 prober can be controlled. (For edit of WIN.INI see below)
    Theta can be memorized and the GPIB switch box is supported.
  - Separate recorded locations supported for OPEN, LOAD and SHORT standards
    Applies to using a custom ISS with SOLT, SOLR, LRM, LRRM or Probe Test
    N/A for TRL. 
  - Added parameters to the Automation method MoveToPos in order
    to support moving to separated port 1 and 2 locations for standards.
  - One port Touchstone S1P type files can now be saved and loaded back
  - Stopping on VNA errors can be turned off completely for 8510 and 87xx
    Edit drvna.ini and in section [Configuration] add/change IgnoreVnaErrMsg=1
  - Added support for HPC probes
  - Added display of load and system impedance values in Probe Test graph

  Enhancements and corrections
  ----------------------------
  - Save and Load of 2-port S-para files are not always defaulting to MDIF
    files. Now last used file format is used as default (.MDF or .S2P)
  - Corrected problems with Anritsu 360 sensing if calibration is ON
  - Corrected problems with using the GPIB switch box with an Anritsu 37xxx VNA
  - Correction to wincal.ini to indicate 7 standard ISS, which adds 114-456
  - Adjusted to new Anritsu 37xxx firmware and allowed returned ID to say
    "Anritsu" instead of "Wiltron".  Same name change in VNA list
  - Fixed a problem with reading switching terms on Anritsu 360 VNA's.  
    Improves LRRM, SOLR and TRL calibrations.  The ratios are read in 
    a way that better relates to actual measurements and gets rid of 
    a small ripple that was present in earlier versions.
  - Fixed a bug where saving S-parameters to a full disk was not detected.
  - Anritsu 360: Fixed problem with getting out of "single sweep and hold mode"
  - Fixed a bug regarding large timeout constants. This corrects some VNA 
    timeout problems for long sweeps (>= 300s), especially for the 360
  - Changed to using no decimals in VNA timeout input.  Was confusing to edit.
  - Fixed a bug that crashed WinCal if a custom ISS, that TRL was last using, 
    was deleted and then the TRL dialog was brought up again.
  - Made a correction to the PPR algorithm if both open and shorted dummy 
    pads are used.
  - Fixed a problem with the GPIB settings for "becoming talker/listener". 
    Especially the PS21 was affected.  It made it hard to reconnect an 
    external host after having been in control of the bus.
  - The debug file "issdata.mdf" is no longer generated during probe test 
    compute
  - Added buttons in config-options-GPIB to manually connect/disconnect the 
    external host with the GPIB switch box.  Mostly for PS21, that has no such 
    button.
  - OLE Automation example is now in Visual Basic 6.0 (was 4.0)
    Updated Visual Basic example uses the VB5/VB6 MSFlexGrid control instead of 
    now obsolete grid32.ocx that was shipped with VB4.  Installable images and 
    the VB 4 project are available on our FTP site ftp.cmicro.com
  - Changed Modify ISS dialog to tabbed property sheet to avoid need for 
    large amounts of DOS type memory.  Also uses less screen area.
    Added check box to custom ISS dialog to disable need to confirm each 
    location
  - Avoided use of void return type in OLE Automation functions.  
    LabView has a problem with the way Borland implemented that.
    They now return bool. (True = success and some are always true)
  - Fixed problems with deciding which standards are needed in LRM/LRRM
    and TRL.  Did not correctly update when changing between OPEN port 1 or 2 
    for stability
  - Fixed F1 help for tabbed dialogs in Windows 95/98
  - Fixed a bug that prevented using 0.0 dB for thru delay in LRM/LRRM
  - Fixed a bug that prevented using 0.0 fF for C-open in probe spec
  - Fixed help access for cal setup dialog.  Was getting main help
  - Fixed handling of user aborted calibrations. The ESC key now works
  - The F1 key now works in all the dialogs that have associated help
  - Fixed problem with unwanted SRQ from HP8510 after the user aborted 
    a measurement
  - Improved reading back binary raw cal files. Some data files had problems,
    but are now readable.  Had to do with end of file condition
  - Adjusted how the switching terms are read to support 8510FX
  - HP 8700 series will only display last VNA error, not the whole error queue
  - Better error messages if file operations fail